Philips North America Llc recall (Class II): Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation. Product: Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Published by source
2026-10-07
First seen by Reality Radar
2026-10-20
Last verified
2026-10-20

Affected

Manufacturer
Philips North America Llc
Products
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Whether your exact unit is affected depends on the model, serial number and region in the official notice.

What the official notice says

Risk: Class II: The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.